| AT WN | [Automatic particle selection] |
| AT SA | [Automatic particle picking - statistical parameters ||] |
| AT IT | [Automatic particle picking - intermediate program.] |
.NEIGHBORHOODPOINTS PIXELS FOR SEARCH: 13
[Enter the number of pixels which corresponds approximately to the size
of the particle. The filtered image is searched for peaks over a region
equal to 13 pixels in this case i.e., the highest peak in a 13 by 13 square
is considered. This way picking up the same particle is avoided and also
nearest neighbour distance is taken into account. Must be 3 or greater.]
.THRESHOLD FOR PEAKS: 0.7
[Enter the threshold value i.e., only peaks having correlation coefficients
over this value are picked up.]
.MICROGRAPH EDGE DIMENSIONS: 80,80
[Peaks closer to the edge of the micrograph are excluded. i.e.,in the example
if x1and y1 are the x- and y- corrdinates of the peaks, and ns,nr,nsam
and nrow are the dimensions of the micrograph edge and the micrograph
respectively then the following conditions are met:
(x1-(ns/2+1)) is greater than or equal to 1.
(x1+(ns/2+1)) is lesser than or equal to nsam.
(y1-(nr/2+1)) is greater than or equal to 1.
(y1+(nr/2+1)) is lesser than or equal to nrow.
.DOCUMENT FILE: DOC001
[Enter the name of the document file where you want the peaks to be
stored.]
NOTES